Laboratory analyst questions and answers.
Pharmaceutical Testing
laboratory participated in interlaboratory comparisons over the last three
years for the analysis additive B in amoxicillin tablets with bias results of
2.4; 2.7;1.9;1.4;1.8 and 2.9. During that mentioned comparisons, the total
number of participated laboratories were 34 with the mean standard deviation of
8.8. During the routine sample analysis, the laboratory has been conducting
quality control with the control limit that was set to +/- 3.34% at 95%
confidence interval. You are required to estimate the expanded uncertainty
associated to the analysis of additive B (K=2).
Considering that the
customer demand on expanded uncertainty is +/- 10%. What would be your
conclusion on the estimated expanded uncertainty?